LSCS-3/LSCS-6
Lshun
Main functions and features:
1. The JKCS series automatic acid value testers, available in three-cup and six-cup models, are fully automated instruments that employ the extraction method to accurately determine the acid value of insulating oils. While enhancing work efficiency and testing accuracy, these instruments also minimize operators’ exposure to samples and reagents, thereby maximizing their personal safety.
2. The extra-large color touch LCD screen is easy to operate via a touch-sensitive LCD display.
3. All the reagents provided with this series of instruments are solid. Users do not need any specialized training or special equipment—simply follow the instructions and use commonly available reagents purchased from local chemical supply stores to prepare solutions on demand. The method is straightforward, the operation is easy, and the testing costs are significantly reduced. Moreover, it eliminates the many inconveniences associated with the long-distance transportation of hazardous chemicals.
4. During testing, simply inject the sample into the sample cup, place the sample cup into the designated cup slot, and the instrument will automatically perform the following operations: sample detection, injection of the extraction solvent and blank value titration, testing, waste liquid discharge, automatic cleaning and shutdown, as well as display and printing of results. A single start-up allows testing of 3 to 6 samples at a time, making it convenient to use and highly efficient.
5. The instrument is equipped with standard acids for calibration and a calibration procedure, enabling users to calibrate both the instrument and the neutralizing solution at any time, thereby overcoming the drawback that the neutralizing solution causes changes in the mid-concentration range. Meanwhile, during the test, the instrument automatically subtracts the background value of the extraction solvent, eliminating the impact of variations in the background value during the use of the extraction solvent on the results. This ensures more accurate and reliable test results.
Main technical specifications:
1. Test range: 0.002–1 mg KOH/g
2. Measurement Error:≤±0.005mg KOH/g
3. Repeatability: ≤ 0.005 mg KOH/g
4. Applicable temperature: 10℃~45℃
5. Applicable humidity: ≤ 85% RH
6. Power supply: AC 220V/50Hz
7. Power: 50W
VRP Series Varnish Removal Oil Purifier
Application
Used in gas and steam turbine, compressor, to remove dissolved and suspended soft pollutants in oil products, sludge and other harmful substances. Avoid turbine failures and expensive oil changes due to paint film problems. When the film is formed, the loss of production is very high. The initial formation of the film is known as a soft pollutant, caused by hot spots in the system, such as bearings, pumps, and high flow on-line filtration. Recent studies have found that the existence of soft pollutants can be divided into dissolved state and suspended state, by removing these soft oxides can avoid the production of paint film. Once the film forms, it will clog valves, filters and other small links, and the life of the oil will be significantly reduced.
The formation of cleaning film will cause the following adverse consequences: valve adhesion, loss of control, resulting in unit failure or startup failure; Filter plugging, limiting oil flow, resulting in oil temperature rise and wear increase; Heat exchange failure, oil temperature rise; Sandpaper surface, increase component wear; Forming paint on the bearing, limiting flow, increasing wear and temperature; Frequent oil changes and system flushes. When the soft pollutants are dissolved in the oil, typically when the oil temperature is above 40℃, they cannot be removed by ordinary mechanical filtration or electrostatic filtration, and these soft pollutants show natural magnetism. For polar absorption, the cooler metal surface "cold spot", i.e. the valve and cooler. When the temperature of the oil is lowered, the thermal stability of the soft pollutants is less than that of the oil, so they are more likely to bake hard on hot surfaces, such as axial shafts.
VRP paint film cleaner. It achieves revolutionary and efficient removal of soft pollutants (dissolved and suspended) from oil products, including gas and steam turbines and compressors under high temperature operation. The warm oil is pumped from the lowest point at the bottom of the tank to the VRP paint film purifier by the equipment's own transfer pump. After cooling treatment (using the thermal stability of the paint film is worse than that of oil), the paint film, particulate matter and moisture are removed by the filter element with wood fiber as the raw material (no need to use ion-exchange resin filter material, greatly saving the cost of use).
Features:
● Increase system reliability and stability. ● No more shutdowns and startup failures caused by paint film.
● No longer need to clean the tank and system flushing. ● Increase jacking oil pressure, more stable bearing temperature.
● Improve the life of oil products, blood additives and system components, such as bearings, valves, seals, etc.
● Great savings can be achieved by avoiding turbine failure and extending the life of oil products. Features:
● Increase system reliability and stability. ● No more shutdowns and startup failures caused by paint film.
● No longer need to clean the tank and system flushing. ● Increase jacking oil pressure, more stable bearing temperature.
● Improve the life of oil products, blood additives and system components, such as bearings, valves, seals, etc.
● Great savings can be achieved by avoiding turbine failure and extending the life of oil products.
Technical Specification
|
Model |
VRP-10 |
|
Flow |
10L/min |
|
Working Pressure |
≤0.5Mpa |
|
Total Power |
5.55KW |
|
Power Supply |
AC380V/50Hz |
|
Overall Dimension |
1466mm×1126mm×1750mm |
|
Net Weight |
500kg |
|
Particle contamination degree(NAS1638)
|
≤6grade |
|
MPC data |
<15 |
■Above size and weight of this equipment are for reference only, the specific data shall be subject to its physical object.

why removing oil varnish?
The causes of varnish in industrial oil systems. We looked at how oil degradation produces submicron that eventually cluster and stick to metal surfaces, turning into varnish. Left unchecked, varnish can lead to varying levels of equipment failure, costly downtime, expensive repairs or, in the worst cases, complete replacement.
When the lubricant inevitably oxidizes and varnish precursors are formed, varnish removal systems are necessary to prevent degradation products from accumulating to the point where varnishing occurs.
Why be concerned about varnish?
Varnish contamination has become a significant maintenance problem in many critical hydraulic and lubrication oil applications, such as gas turbines, compressors, and hydraulic presses. Varnish contamination in Gas Turbine Lube Oil Systems has been a particularly high profile issue over the last decade. A very small amount of varnish that plates out on the metal surfaces within the lube oil system (such as a servo-valve) can cause inefficient operation or prevent the system from starting properly, resulting in a fail-to-start condition or in some cases a unit trip.
how can it be prevented or eliminated?
Our Varnish Removal Systems (VRP) were developed to address the need to properly remove varnish (also known as lacquer, sludge, or tar) commonly found in various lubrication and hydraulic systems. Our process utilizes either granular adsorbent media or depth media filter elements to effectively remove dissolved and suspended oil degradation products (i.e. the cause of varnish). Varnish removal is accomplished with oxidatively stable media, engineered to selectively remove only the varnish-causing contaminants without disturbing the oil's additive package.